FDANovember 8, 2019device

CrossCath Support Catheter, RPN CXC3.4-2.2-18-150-P-NS-0, Order Number G51577 For diagnostic and interventional vascular procedures.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

A manufacturing error may cause the radiopaque marker bands to be too loose on certain CXC3.0 CrossCath Support Catheters (compatible with 0.014 wire guides) and too tight on certain CXC3.4 CrossCath Support Catheters (compatible with 0.018 wire guides).

What to do

FDA enforcement status: Terminated

Brands named

cook

UPCs

00827002515772