FDANovember 7, 2019device

iChem VELOCITY Urine Chemistry Analyzer

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Beckman Coulter has become aware of an issue affecting Specific Gravity chemistry settings for the iChemVELOCITY Urine Chemistry System that prevents the trigger of a chemistry system flag S3 and may result in erroneous Specific Gravity (SG) results.

What to do

FDA enforcement status: Terminated

Brands named

beckman coulterbeckman