FDAOctober 21, 2019device

MicroScan WalkAway-40 plus Reconditioned Instrument REF B1018-340R

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere

What to do

FDA enforcement status: Ongoing

Brands named

beckman coulterbeckman

UPCs

15099590684112