FDANovember 13, 2024device

DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The DxI 9000 Access Immunoassay Analyzer wash wheel mixer may not spin freely within the idler pulley housing. When the problem occurs, the analyzer may generate a system event for a Wash Wheel Dispense Arm Spin Mixer Error that is displayed on the navigation bar in the User Interface. The analyzer enters the red system status and will generate a SYS flag. The tests that are in progress may be cancelled and may cause a delay in reporting patient results

What to do

FDA enforcement status: Ongoing

Brands named

beckman coulterbeckman