FDASeptember 4, 2019device

MicroScan WalkAway-40 plus Reconditioned Instrument

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Instruments were manufactured with a shortened 30mL reagent draw straw which may result in reagents not dispensing and may lead to organism misidentification and/or Quality Control (QC) failure.

What to do

FDA enforcement status: Terminated

Brands named

beckman coulterbeckman

UPCs

15099590684112