FDAFebruary 5, 2025device

Access 2 Immunoassay Analyzer, Catalog Number 81600N

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Beckman Coulter identified that an issue for Access 2 Sample Motors (B48016) due to improper alignment, causing the motors to rub against the sample wheelbase plate and index sensor. The issue was traced to a worn tool used in production (at the supplier), affecting Access 2 sample motors. This could lead to a potential delay in reporting patient test results.

What to do

FDA enforcement status: Ongoing

Brands named

beckman coulterbeckman

UPCs

15099590265380

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.