FDAAugust 15, 2019device

Total Bilirubin Reagent (TBIL 2 x 300 ) and (TBIL 2 x 400), Ref:442745 (300 tests/cartridge) and 476861 (400 tests/cartridge), For In Vitro Diagnostic Use, Rx Only, UDI: 15099590233075 (300 tests/cartridge) and 15099590625597 (400 tests/cartridge)

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

It has been determined that N-acetyl p benzoquinone imine (NAPQI), a metabolite of acetaminophen (paracetamol), may cause negative interference for assays if present in high quantities in serum due to acetaminophen overdoese. This could result in a delay in diagnosis and treatment.

What to do

FDA enforcement status: Terminated

Brands named

beckman coulterbeckman

UPCs

1509959023307515099590625597