FDAAugust 15, 2019device
Total Bilirubin Reagent (TBIL 2 x 300 ) and (TBIL 2 x 400), Ref:442745 (300 tests/cartridge) and 476861 (400 tests/cartridge), For In Vitro Diagnostic Use, Rx Only, UDI: 15099590233075 (300 tests/cartridge) and 15099590625597 (400 tests/cartridge)
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
It has been determined that N-acetyl p benzoquinone imine (NAPQI), a metabolite of acetaminophen (paracetamol), may cause negative interference for assays if present in high quantities in serum due to acetaminophen overdoese. This could result in a delay in diagnosis and treatment.
What to do
FDA enforcement status: Terminated
Brands named
beckman coulterbeckman
UPCs
1509959023307515099590625597
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAAccess Total T4 Calibrator, Catalog No. 338052026-03-16
- FDACHEMISTRY ANALYZER AU5800, REF: B96697, B96698, B23279, B23280, B232812026-01-30
- FDADxC 700 AU, REF: B86444, B864462026-01-30
- FDACellMek SPS Sample Preparation System, REF: C44603, with software2025-12-23
- FDAThe LDL-Cholesterol assay is comprised of two distinct phases. In phase one a unique detergent solubilizes cholesterol from non-LDL- lipoprotein particles. This cholesterol is consumed by cholesterol esterase, cholesterol oxidase, peroxidase and 4- aminoantipyrine to generate a colorless end prod...2025-12-08
- FDABeckman Coulter UniCel Dxl 800, Part numbers: 973100, A71456, A71457, A25288; Immunoassay Method, Troponin Subunit2025-11-26
- FDABeckman Coulter UniCel Dxl 600, Part numbers: A71460, A71461, A30260; Immunoassay Method, Troponin Subunit2025-11-26
- FDACOULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conjunction with a cyanide-free lytic agent for counting and sizing blood cells on all UniCel DxH Coulter Cellular Analysis Systems. For use as a rinsing agent on the UniCel DxH Slidemake...2025-11-10