FDAMarch 27, 2025device

Beckman Coulter DxI 9000 Access Immunoassay Analyzer

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

When DxI 9000 Access Immunoassay Analyzer is connected to a host system (laboratory information system (LIS) or middleware) and has accumulated canceled QC test results that were not sent by LIS, the analyzer may lose communication with the host system. The lost connection interrupts sample processing, which delays reporting patient test results and may subsequently delay patient treatment.