FDAJanuary 17, 2020device

LabPro Data Management System, UDI: 15099590676872 - Product Usage: The LabPro Data Management System is a Microsoft Windows based software program and is intended to manage both microbial identification (ID) and antimicrobial agent susceptibility testing (AST) data generated from MicroScan instr...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

A security-only update was released via remote diagnostics on January 14, 2020 to customers with Windows 7 operation system. Shortly after the releases some customers reported that Data Management System computers would not restart. If the computer is unable to restart, then the system cannot download data generated from MicroScan instruments or manually entered microbiology test results, resulting in a potential of delayed results.

What to do

FDA enforcement status: Terminated

Brands named

beckman coulterbeckman

UPCs

15099590676872

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
LabPro Data Management System, UDI: 15099590676872 - Product Usage: The LabPro Data Management System is a Microsoft Windows based software program and is intended to manage both microbial identification (ID) and antimicrobial agent susceptibility testing (AST) data generated from MicroScan instr... — Recall Details · AllClear