FDAAugust 16, 2022device

Bentson Plus Wire Guide (Extra Body/20cm Distal Flexibility with Softened Tip), Guidewire; Reference Part Number: TSFBP-35-180 (GPN G09782) TSFBP-35-145 (GPN G09735)

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.

What to do

FDA enforcement status: Ongoing

Brands named

cook

UPCs

0082700209782700827002097353