FDAApril 6, 2016device

EZ Slide Battery/line powered hydraulic table with removable leg section and 21 top slide General Purpose Surgical Table Skytron EZ Slide Operating Table General Purpose Surgical Table

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The potential exists within the identified tables that incorrectly manufactured parts have been used in assembly. If a table is found to have these incorrectly manufactured parts, it is possible for the tables elevation column slide cap to become detached from the elevation column and could result in the loss of control of the table top and injury to the patient or staff.

What to do

FDA enforcement status: Terminated

Brands named

skytron div the kmwskytronskytron div