FDANovember 16, 2017device

ALM XTEN Surgical Light intended to provide visible illumination for the surgical field or for the examination of the patient. Part Numbers: ARD567801093, ARD567801094

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Getinge has received complaints concerning the X TEN Surgical Lights whose spring arms have broken, causing light heads to separate from the assembly. The issue relates to cracking of the spring arms at the weld seams that has the potential to develop into breaks with heavy use and aging of the equipment.

What to do

FDA enforcement status: Terminated

Brands named

getinge us salesgetingegetinge us