FDAApril 13, 2016device

Cholesterol, Catalog No. OSR6x16 Product Usage:Cholesterol OSR6x16 is intended as a System reagent for the quantitative determination of Cholesterol concentrations in human serum on Beckman Coulter AU analyzers.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Beckman Coulter has identified that Dipyrone (Metamizole) may cause erroneous, false low results in the following assays: Cholesterol OSR6X16, Triglyceride OSR6X118, Uric Acid OSR6X98, Lactate OSR6X93, and Lipase OSR6X30.

What to do

FDA enforcement status: Terminated

Brands named

beckman coulterbeckman