FDAJune 20, 2023device

Ringed DxI Reaction Vessels (RVs)

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Beckman Coulter, Inc. is recalling Ringed DxI Reaction Vessels (RVs) as part of Unicel DxI Immunoassay System (600, 800 series). The reason for the recall is that the firm has identified 3 lots of RVs that failed an internally used service test to evaluate washing efficiency. Increased relative light units (RLU) suspected to be caused by the resulting rough interior surface may lead to artificially falsely elevated results for susceptible tests, in particular, sandwich assays including, but not limited to, hsTnI, Hybritech p2PSA, IPTH/IPTH IO.

What to do

FDA enforcement status: Ongoing

Brands named

beckman coulterbeckman

UPCs

15099590229979

Recall history