FDAFebruary 29, 2024device

Beckman Coulter DxI 9000 Access Immunoassay Analyzer is a discrete photometric chemistry analyzer for clinical use. It is an in-vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids with Cata...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The reason for the recall is the pneumatic tubing used in the analytical pick-and-place (PnP) assembly cable may fully or partially disconnect from the flex cable assembly fittings during normal operation. The DxI 9000 Access Immunoassay Analyzer may generate a System Event log message that indicates PnP errors. Any cancelled tests will have an associated SYS flag. The analyzer may enter the Red-system status because of this issue. The analyzer also may not eject RVs, where unbound conjugate could potentially splash up on the sidewall of the RV, not being properly washed away, resulting in an artificially higher signal (RLU). The analyzer may not detect this splashing and continue with normal operation, which can cause false low or high erroneous results. Consequently, the analyzer may cancel in-progress tests. These errors could potentially cause delay reporting patient test results and patient treatments. The degree of the potential health hazard is dependent on the analytes that are

What to do

FDA enforcement status: Ongoing

Brands named

beckman coulterbeckman

UPCs

15099590732103

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →