FDAApril 30, 2019device
Beckman Coulter ACCESS Immunoassay Systems, Access Thyroglobulin Antibody II, REF A32898 (UDI: 15099590202910) for use with the Access Family of Immunoassay Systems including: Access 2, UniCel DxI 600, UniCel DxI 800, UniCel DxC 600i, UniCel DxC 660i, UniCel DxC 680i, UniCel DxC 860i, and UniCel ...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
The recalling firm characterized the potential interference effects of biotin (up to 1,200 ng/mL) on all biotin-mediated Access assays. The following six assays: Free T3-REF A13422, Total T3-REF 33830, Free T4-REF 33880, GI Mon-387687, Tg-REF 33860, and TgAbII-REF A32898 displayed significant interference (defined as a result bias greater than +/-10%) at 1,200 ng/mL of biotin.
What to do
FDA enforcement status: Terminated
Brands named
beckman coulterbeckman
UPCs
15099590202910
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAAccess Total T4 Calibrator, Catalog No. 338052026-03-16
- FDACHEMISTRY ANALYZER AU5800, REF: B96697, B96698, B23279, B23280, B232812026-01-30
- FDADxC 700 AU, REF: B86444, B864462026-01-30
- FDACellMek SPS Sample Preparation System, REF: C44603, with software2025-12-23
- FDAThe LDL-Cholesterol assay is comprised of two distinct phases. In phase one a unique detergent solubilizes cholesterol from non-LDL- lipoprotein particles. This cholesterol is consumed by cholesterol esterase, cholesterol oxidase, peroxidase and 4- aminoantipyrine to generate a colorless end prod...2025-12-08
- FDABeckman Coulter UniCel Dxl 800, Part numbers: 973100, A71456, A71457, A25288; Immunoassay Method, Troponin Subunit2025-11-26
- FDABeckman Coulter UniCel Dxl 600, Part numbers: A71460, A71461, A30260; Immunoassay Method, Troponin Subunit2025-11-26
- FDACOULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conjunction with a cyanide-free lytic agent for counting and sizing blood cells on all UniCel DxH Coulter Cellular Analysis Systems. For use as a rinsing agent on the UniCel DxH Slidemake...2025-11-10