UIBC (Unsaturated Iron Binding Capacity), REF: OSR61205,
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Beckman Coulter identified that the Unsaturated Iron Binding Capacity (UIBC) assay is not meeting their labeled hemoglobin interference claims of no significant interference (i.e., less than 10%) up to 200 mg/dL hemoglobin as indicated in their instructions for use. Initial internal testing confirmed that samples with low UIBC levels failed the claimed hemolysis interference specifications with reported biases up to -43.6 % UIBC when hemolyzed serum samples contained 200 mg/dL hemoglobin.
What to do
FDA enforcement status: Ongoing
Brands named
UPCs
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAAccess Total T4 Calibrator, Catalog No. 338052026-03-16
- FDACHEMISTRY ANALYZER AU5800, REF: B96697, B96698, B23279, B23280, B232812026-01-30
- FDADxC 700 AU, REF: B86444, B864462026-01-30
- FDACellMek SPS Sample Preparation System, REF: C44603, with software2025-12-23
- FDAThe LDL-Cholesterol assay is comprised of two distinct phases. In phase one a unique detergent solubilizes cholesterol from non-LDL- lipoprotein particles. This cholesterol is consumed by cholesterol esterase, cholesterol oxidase, peroxidase and 4- aminoantipyrine to generate a colorless end prod...2025-12-08
- FDABeckman Coulter UniCel Dxl 800, Part numbers: 973100, A71456, A71457, A25288; Immunoassay Method, Troponin Subunit2025-11-26
- FDABeckman Coulter UniCel Dxl 600, Part numbers: A71460, A71461, A30260; Immunoassay Method, Troponin Subunit2025-11-26
- FDACOULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conjunction with a cyanide-free lytic agent for counting and sizing blood cells on all UniCel DxH Coulter Cellular Analysis Systems. For use as a rinsing agent on the UniCel DxH Slidemake...2025-11-10
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.
Start free — 200 items, no card →