FDAJune 25, 2025device

DxI 800 Access Immunoassay Analyzer W/Spot B, Part Number A71456

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed.

What to do

FDA enforcement status: Ongoing

Brands named

beckman coulterbeckman