FDAMay 15, 2017device

Beckman Coulter Synchron AQUA CAL Level 1, Part No. 471288 Product Usage: Used for the system calibration of Na, K, Cl, Urea N, Urea, Glucose Creatine, Ca, CO2, and P in UniCel DxC Systems.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The firm identified a negative shift of 5mmol/L for sodium recovery in Synchron Control Level 3 when calibrated with AQUA Cal 1.

What to do

FDA enforcement status: Terminated

Brands named

beckman coulterbeckman