FDASeptember 18, 2018device

The Access Cortisol Reagent. REF (Catalog Number) 33600, Lot 831755. A paramagnetic particle, chemiluminscent immunoassay. Product Usage: The assay is a paramagnetic particle, chemiluminescent immunoassay for the quanitative determination of cortisol levels in human serum, plasma (heparin, EDTA) ...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Cross contamination may have occurred between wells for the reagent pack. As a result of this defect, patient results may be negatively impacted (falsely decreased) by up to 16%).

What to do

FDA enforcement status: Terminated

Brands named

beckman coulterbeckman

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.